Hill-Rom TotalCare Duo 2 를 위한 안전성 경고 / 현장 안전성 서한

Medicines and Healthcare products Regulatory Agency에 따르면, 해당 안전성 경고 / 현장 안전성 서한 는 United Kingdom 에서 manufacturer #17429 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

현장 안전성 서한은 의료기기 제조사 또는 제조사 대리인이 시중에 유통되고 있는 자사 의료기기 제품에 대하여 취할 수 있는 조치를 알리는 소통방식입니다. 주요 대상층은 보건의료업계 종사자와 의료기기 사용자들입니다. 여기에는 회수(recall) 조치와 경고도 포함될 수 있습니다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Safety alert / Field Safety Notice
  • 사례 ID
    MDA/2007/065
  • 사례 연번
    CON2032088
  • 날짜
    2007-07-22
  • 사례 국가
  • 사례 출처
    MHRA
  • 사례 출처 URL
  • 비고 / 경고
    Data from the United Kingdom is current through April 2019. All of the data comes from the Medicines and Healthcare products Regulatory Agency, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and the United Kingdom.
  • 데이터 추가 비고
  • 원인
    The manufacturer has omitted to provide suitable decontamination instructions with this pressure relieving hospital bed. the mhra has received two reports of damaged mattress covers that allowed body fluids to soak into the mattress. no instruction was provided on how to check the condition of the mattress and cover. this presents a risk of cross infection. the manufacturer has omitted to provide customers with information on how to undertake suitable checks and decontamination procedures. the manufacturer does not include full decontamination details with the instructions for use, due to different requirements in different countries. in the uk they are covered in a separate document, w13001uks rev2. two nhs trusts have reported a risk of cross infection with this bed. residues of body fluids from previous occupants have been found under mattress covers. many of these beds are used by intensive care units. the mhra has not received any adverse incident reports of cross infection occurring due to this problem.
  • 조치
    Trace all Hill-Rom TotalCare Duo 2 beds in use and in storage. Contact Hill-Rom on 01530 562167 for copies of the cleaning and disinfection instructions. Unzip the mattress covers and check for signs of fluid contamination on the mattress and damage or holes in the mattress covers. Dispose of damaged covers. If replacement or spare mattresses and mattress covers are required, contact Hill-Rom on 01530 411000 Between each patient use check whether any fluids have penetrated the cover and follow Hill-Rom’s cleaning, disinfection and laundry advice for the covers and mattresses.