Medical oxygen and medical air cylinders filled to 200 atmospheres (200 bar/20,000 kPa) 를 위한 안전성 경고 / 현장 안전성 서한

Medicines and Healthcare products Regulatory Agency에 따르면, 해당 안전성 경고 / 현장 안전성 서한 는 United Kingdom 에서 manufacturer #17429 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

현장 안전성 서한은 의료기기 제조사 또는 제조사 대리인이 시중에 유통되고 있는 자사 의료기기 제품에 대하여 취할 수 있는 조치를 알리는 소통방식입니다. 주요 대상층은 보건의료업계 종사자와 의료기기 사용자들입니다. 여기에는 회수(recall) 조치와 경고도 포함될 수 있습니다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Safety alert / Field Safety Notice
  • 사례 ID
    MDA/2013/002
  • 사례 연번
    CON224427
  • 사례 출판 날짜
    2013-01-14
  • 사례 국가
  • 사례 출처
    MHRA
  • 사례 출처 URL
  • 비고 / 경고
    Data from the United Kingdom is current through April 2019. All of the data comes from the Medicines and Healthcare products Regulatory Agency, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and the United Kingdom.
  • 데이터 추가 비고
  • 원인
    Risk of serious injury to staff and patients. currently 137 bar cylinders are widely used in the uk. connection to significantly higher pressure 200 bar cylinders may cause damage to devices such as anaesthetic machines, ventilators, incubators, respiratory support and resuscitation equipment. a hospital trust is undertaking a pilot scheme to introduce 200 bar cylinders into clinical use. while the trust involved has ensured the safety of these cylinders within their organisation, there are concerns that these cylinders could be moved out of the trust during patient discharge, or patient transfer in ambulances for use with transport incubators and transport ventilators.
  • 조치
    Be aware that 200 bar medical oxygen and medical air cylinders will be entering clinical use in a pilot scheme.  If this pilot is successful, the use of 200 bar cylinders is expected to become widespread across the UK. Before connecting these 200 bar cylinders to medical devices staff must confirm that it is safe to operate at this higher pressure by checking the instructions for use, or contacting the manufacturer. All users should check the cylinder label before use to ensure the cylinder gas and pressure are appropriate and that the expiry date has not been exceeded. Staff must seek permission to remove cylinders for use outside the hospital and ensure the above checks are carried out prior to use.

Device

  • 모델명 / 제조번호(시리얼번호)
  • 제품 설명
    Medical oxygen and medical air cylinders filled to 200 atmospheres (200 bar/20,000 kPa). Various sizes and outlets. Supplied by SOL SpA. Medical oxygen at 200 bar with pin index valve in 2, 5 and 50 litre capacity cylinders. Medical oxygen at 200 bar with bull nose valve in 10 and 20 litre capacity cylinders. Medical air at 200 bar with pin index valve in 2, 5, 10, 20 and 50 litre capacity cylinders.