Primus anaesthetic workstation 를 위한 안전성 경고 / 현장 안전성 서한

Medicines and Healthcare products Regulatory Agency에 따르면, 해당 안전성 경고 / 현장 안전성 서한 는 United Kingdom 에서 manufacturer #17429 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

현장 안전성 서한은 의료기기 제조사 또는 제조사 대리인이 시중에 유통되고 있는 자사 의료기기 제품에 대하여 취할 수 있는 조치를 알리는 소통방식입니다. 주요 대상층은 보건의료업계 종사자와 의료기기 사용자들입니다. 여기에는 회수(recall) 조치와 경고도 포함될 수 있습니다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Safety alert / Field Safety Notice
  • 사례 ID
    MDA/2007/040
  • 사례 연번
    CON2031245
  • 날짜
    2007-05-23
  • 사례 국가
  • 사례 출처
    MHRA
  • 사례 출처 URL
  • 비고 / 경고
    Data from the United Kingdom is current through April 2019. All of the data comes from the Medicines and Healthcare products Regulatory Agency, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and the United Kingdom.
  • 데이터 추가 비고
  • 원인
    Potential misconnection of the backup oxygen cylinder; the oxygen cylinder inlet is situated alongside an oxygen outlet and both are fitted with the same connector.  a misconnection to the oxygen outlet would prevent backup oxygen supply in the event of oxygen pipeline failure. the use of an oxygen backup cylinder is optional on the primus.  if the backup cylinder is required the user must attach the cylinder, connect its oxygen nist fitting and connect the gas pressure sensor before use.  the primus has the further user option of an oxygen outlet, which is slightly longer and situated alongside the oxygen cylinder inlet fitting at the back of the primus.  this oxygen outlet is the same nist fitting as the oxygen inlet beside it and therefore users can connect the oxygen cylinder to the wrong fitting. the mhra has received a report of this misconnection, which was not detected in the device’s automated pre-use checks or the pre-use checks recommended by the aagbi (association of anaesthetists of great britain and ireland).  the oxygen pressure gauge indicated that the backup oxygen cylinder was full so the misconnection was not apparent to the user.
  • 조치
    Ensure the backup oxygen cylinder’s NIST (non-interchangeable screw threaded) fitting is connected to the appropriate oxygen inlet.  The correct connection guide can be found above the gas inlet block at the back of the device. Users are recommended that pre-use checks should include the disconnection of the oxygen pipeline to ensure the continuity of oxygen supply from the backup cylinder.

Device

  • 모델명 / 제조번호(시리얼번호)
  • 제품 설명
    Primus anaesthetic workstation supplied by Draeger Medical UK Ltd. The Primus anaesthetic workstation is a modern machine with an integrated ventilator and gas monitoring system for the administration of anaesthesia to adults, children and neonates.