Toshiba ultrasound transducers 를 위한 안전성 경고 / 현장 안전성 서한

Medicines and Healthcare products Regulatory Agency에 따르면, 해당 안전성 경고 / 현장 안전성 서한 는 United Kingdom 에서 manufacturer #17429 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

현장 안전성 서한은 의료기기 제조사 또는 제조사 대리인이 시중에 유통되고 있는 자사 의료기기 제품에 대하여 취할 수 있는 조치를 알리는 소통방식입니다. 주요 대상층은 보건의료업계 종사자와 의료기기 사용자들입니다. 여기에는 회수(recall) 조치와 경고도 포함될 수 있습니다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Safety alert / Field Safety Notice
  • 사례 ID
    MDA/2014/046
  • 사례 연번
    CON484998
  • 날짜
    2014-12-08
  • 사례 출판 날짜
    2014-12-08
  • 사례 국가
  • 사례 출처
    MHRA
  • 사례 출처 URL
  • 비고 / 경고
    Data from the United Kingdom is current through April 2019. All of the data comes from the Medicines and Healthcare products Regulatory Agency, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and the United Kingdom.
  • 데이터 추가 비고
  • 원인
    Counterfeit ultrasound transducers have been supplied to 2 hospitals. these devices may not meet the performance specified by toshiba. there have been no reports of patient harm. transducers supplied directly by toshiba medical systems uk are not affected. mhra and toshiba will continue to investigate this matter as new cases appear.   the mhra has become aware that counterfeit toshiba ultrasound transducers have been found in two hospitals in the uk. as there is no evidence of patient harm, mhra is not recommending that patients are called back to their hospital or clinic. we are publishing this alert as a precautionary measure to raise awareness of the issue and to ask users to check toshiba ultrasound transducers not supplied direct from toshiba medical systems uk. as counterfeits may look very similar to legitimate products, they need to be checked very carefully.  however, a competent user should easily identify a fake transducer. in addition, the device should have undergone quality checks before being brought into use. these devices are also subject to periodic quality and maintenance checks. there is guidance for healthcare providers from ipem, sor and cqc on the procurement, initial acceptance testing, quality assurance and maintenance of these devices. this would mean that counterfeits should be identified before use.
  • 조치
    Identify and quarantine any suspect devices. Further advice on how to identify fake devices is given on pages 2 and 3 of this alert. To authenticate suspected devices, contact your local Toshiba representative via their dedicated email report.counterfeit.uk@tmse.nl Report any counterfeit devices to the MHRA via Yellow Card Counterfeit transducers can be identified by one or more of the following characteristics: The label may be poorly attached and made of paper. It may also peel off easily and leave a residue. Toshiba labels are plastic or metal and are difficult to remove The descriptive text may be incorrect – a linear transducer array may be described as ‘convex’ Since 2005 Toshiba serial numbers are ten digits long e.g. TDA1293472. Any label with fewer numbers is likely to be counterfeit Scanning the QR code (if any) on the label will not take the user to a website or it gives a different probe type and serial number The Toshiba logo or wording may not be as sharp, and the casing may have a different finish when compared to a genuine probe Counterfeit Toshiba transducer heads can be substantially lighter in weight than genuine Toshiba items The pictures below are examples only to help identify suspected counterfeit probes. They do not represent all possible fake devices.

Device

  • 모델명 / 제조번호(시리얼번호)
  • 제품 설명
    Toshiba ultrasound transducers All models