NavLock Tracker 의 안전성 경고

Department of Health에 따르면, 해당 안전성 경고 는 Hong Kong 에서 Medtronic 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

안전성 경고는 의료기기 제품과 관련된 중요한 정보와 권고사항을 담고 있습니다. 물론 안전성 경고가 배포되었다고 해서 해당 제품이 무조건 안전하지 않은 제품이라는 것은 아닙니다. 보건의료업계 종사자와 의료기기 사용자들에게 배포되는 안전성 경고에는 회수(recall)도 포함될 수 있습니다. 제조사가 안전성 경고를 작성하기도 하지만, 보건당국에서 작성하는 경우도 있습니다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Safety alert
  • 날짜
    2017-05-04
  • 사례 출판 날짜
    2017-05-04
  • 사례 국가
  • 사례 출처
    DH
  • 사례 출처 URL
  • 비고 / 경고
    Hong Kong data is current through September 2018. All of the data comes from the Department of Health (Hong Kong), except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of data from the U.S. and Hong Kong.
  • 데이터 추가 비고
    Medical Device Safety Alert
  • 원인
    Medical device safety alert: medtronic navlock tracker the united states food and drug administration (fda) has issued a medical device safety alert concerning navlock tracker, manufactured by medtronic. on 1 may 2017, the manufacturer issued a communication related to the use of non-medtronic instruments with medtronic's navlock tracker. this notice reports that medtronic is aware of occurrences of minor and serious injuries, including two patient deaths, following procedures that involved the use of non-medtronic instruments with medtronic's navlock tracker. as a result, medtronic notified the agency of plans to update its product labeling to clarify the use of non-medtronic instruments with its navlock tracker. per medtronic’s communication, the “indications for use” and "warning" statement found in the labeling of the navlock tracker will be updated to state: revised indications for use: "the navlock trackers are intended to enable navigation of medtronic instrumentation used during spinal fusion and interbody procedures with the medtronic stealthstation surgical navigation system. the navlock trackers should only be used with medtronic instruments." strengthened warning: "the navlock tracker is designed and tested for use only with medtronic instruments. the use of non-medtronic instruments with navlock tracker may result in inaccuracy, leading to serious injury or death." between 1 january 2013 and 22 march 2017, the fda identified a total of 196 medical device reports (mdrs) associated with the use of medtronic’s navlock tracker. however, due to limited information available in the mdrs, it is unclear if, and to what degree, the use of 3rd party instruments caused or contributed to the adverse event. additionally, the fda recognizes that the contributing factors associated with accuracy and precision problems when using stereotaxic navigation systems are multi-factorial. the two deaths noted involved patients undergoing a spinal procedure and experiencing post-operative hemorrhage. these two patient deaths occurred following a procedure that involved the use of 3rd party surgical stereotaxic navigation instruments that have not been cleared by the fda for use with medtronic’s navlock tracker on medtronic’s stealthstation. to date, the fda has reviewed and cleared surgical stereotaxic navigation instruments produced by manufacturers other than medtronic for use with medtronic’s stealthstation. among these are instruments manufactured by alphatec spine, inc., globus medical, inc., and orthofix, inc. which have been cleared for use with medtronic’s navlock tracker on medtronic’s stealthstation. the instruments from these manufacturers (e.G., 3rd party instruments) were cleared based on non-clinical performance data submitted to the fda by those manufacturers. the fda has provided the following recommendations: do not use surgical stereotaxic instruments that have not been cleared by the fda for use with medtronic’s navlock tracker on the medtronic stealthstation surgical navigation system. be aware of medtronic’s communication disclaimer icon related to the use of non-medtronic surgical stereotaxic instruments with medtronic’s navlock tracker. when using any surgical stereotaxic instruments, be aware that surgical accuracy should be assessed repeatedly throughout a procedure when using a surgical navigation system by positioning the navigated instrument tip on an identifiable anatomical landmark and comparing the actual tip location to that displayed by the system. be aware that not all surgical instruments are appropriate for surgical navigation. consider using surgical stereotaxic instruments that are identified as compatible with the surgical navigation system being used. for details, please refer to the fda websites: https://www.Fda.Gov/safety/medwatch/safetyinformation/safetyalertsforhumanmedicalproducts/ucm556432.Htm https://www.Fda.Gov/medicaldevices/safety/letterstohealthcareproviders/ucm556052.Htm if you are in possession of the affected products, please contact your supplier for necessary actions. posted on 4 may 2017.

Device

  • 모델명 / 제조번호(시리얼번호)
  • 제품 설명
    Medical Device Safety Alert: Medtronic NavLock Tracker
  • Manufacturer

Manufacturer

  • 제조사 모회사 (2017)
  • 제조사 의견
    “If our surveillance systems identify a potential performance issue, our personnel promptly evaluate the problem, including, when appropriate, conducting root cause investigations and internal testing to assess whether the product continues to meet specifications and defined performance criteria,” Medtronic told ICIJ in a statement. “In some cases, based on this evaluation, Medtronic may determine that a recall is necessary.” The company said that it communicates with healthcare providers and/or patients and provide recommendations to address such issues. Medtronic noted that these communications can include letters, emails, calls, press releases, physician notifications and social media postings, as well as informing the FDA and other regulators of the actions.
  • Source
    DH