SL75/5 Linear Accelerator 의 안전성 경고

Department of Health에 따르면, 해당 안전성 경고 는 Hong Kong 에서 Elekta 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

안전성 경고는 의료기기 제품과 관련된 중요한 정보와 권고사항을 담고 있습니다. 물론 안전성 경고가 배포되었다고 해서 해당 제품이 무조건 안전하지 않은 제품이라는 것은 아닙니다. 보건의료업계 종사자와 의료기기 사용자들에게 배포되는 안전성 경고에는 회수(recall)도 포함될 수 있습니다. 제조사가 안전성 경고를 작성하기도 하지만, 보건당국에서 작성하는 경우도 있습니다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Safety alert
  • 날짜
    2012-08-29
  • 사례 출판 날짜
    2012-08-29
  • 사례 국가
  • 사례 출처
    DH
  • 사례 출처 URL
  • 비고 / 경고
    Hong Kong data is current through September 2018. All of the data comes from the Department of Health (Hong Kong), except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of data from the U.S. and Hong Kong.
  • 데이터 추가 비고
    Medical Device Safety Alert
  • 원인
    Medical device safety alert: elekta sl75/5 linear accelerator the medicines and healthcare products regulatory agency (mhra), united kingdom posted a field safety notice concerning all sl75/5 linear accelerator, manufactured by elekta. according to the manufacturer, there was malfunction of the semikron diode of sl75/5. this diode is fitted as charging diode d8 but can also be fitted as reverse diode d7 on later machines. they can be found on the modulator chassis assembly. if a malfunction of the semikron diode occurs, the ceramic body of the diode can rupture. this can cause ceramic material and silicone oils to eject from the modulator assembly which may cause injury to on-site service personnel. there are currently no reported incidents of this kind on sl75/5 machines. furthermore, the manufacturer recommends that: service personnel should always wear protective safety glasses, when they do work on the modulator assembly with the covers removed. users should make sure the door of the modulator chassis assembly is closed when the digital accelerator is in operation. for details, please refer to the following mhra website: http://www.Mhra.Gov.Uk/safetyinformation/safetywarningsalertsandrecalls/fieldsafetynotices/index.Htm if you are in possession of the affected product, please contact your supplier for necessary actions. posted on 29 august 2012.

Device

  • 모델명 / 제조번호(시리얼번호)
  • 제품 설명
    Medical Device Safety Alert: Elekta SL75/5 Linear Accelerator
  • Manufacturer

Manufacturer