Accu-Chek Spirit Combo insulin pump (Pumps with serial numbers in the range 10171897 to 10281629 inclusive) 에 대한 현장 안전성 서한

Health Products Regulatory Authority에 따르면, 해당 현장 안전성 서한 는 Ireland 에서 Roche Diagnostics Ltd. 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

현장 안전성 서한은 의료기기 제조사 또는 제조사 대리인이 시중에 유통되고 있는 자사 의료기기 제품에 대하여 취할 수 있는 조치를 알리는 소통방식입니다. 주요 대상층은 보건의료업계 종사자와 의료기기 사용자들입니다. 여기에는 회수(recall) 조치와 경고도 포함될 수 있습니다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Field Safety Notice
  • 사례 연번
    SN2014(13)
  • 날짜
    2014-03-24
  • 사례 국가
  • 사례 출처
    HPRA
  • 사례 출처 URL
  • 비고 / 경고
    "Irish data is current through April 2019. All of the data comes from the Health Products Regulatory Authority (Ireland), except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of data from the U.S. and Ireland."
  • 데이터 추가 비고
  • 원인
    There is an increased risk that, due to a component change, vibration alarm defects may cause an ‘e-7’ alert message. this fault will be detected at pump start up, when it will display an ‘e-7’ error message and give an audible signal, but will fail to start. the other pump alarms – visual and audible - are not affected by this issue. pumps with serial numbers in the range 10171897 to 10281629 (inclusive) are potentially affected. these pumps were distributed to the irish market between late 2010 and early 2012 and to date, no ‘e-7’ issues related to vibration alarm defects have been reported from ireland.
  • 조치
    1. Identify affected pumps.  2. If a delay to insulin therapy could compromise patient safety, consider using an alternative device.  3. When using these pumps:  • Ensure that users are aware of the problem  • Ensure that users have a syringe or insulin pen available in case the pump stops  • If the pump displays an ‘E-7’ error, contact Roche for a replacement device  4. Ensure that appropriate personnel are made aware of this notice. Please also pass this notice on to any end user or organisation where the potentially affected insulin pumps have been transferred

Device

  • 모델명 / 제조번호(시리얼번호)
  • 제품 설명
    Accu-Chek Spirit Combo insulin pump (Pumps with serial numbers in the range 10171897 to 10281629 inclusive)