ARCHITECT CA 19-9XR 에 대한 현장 안전성 서한

Health Products Regulatory Authority에 따르면, 해당 현장 안전성 서한 는 Ireland 에서 Abbott Laboratories 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

현장 안전성 서한은 의료기기 제조사 또는 제조사 대리인이 시중에 유통되고 있는 자사 의료기기 제품에 대하여 취할 수 있는 조치를 알리는 소통방식입니다. 주요 대상층은 보건의료업계 종사자와 의료기기 사용자들입니다. 여기에는 회수(recall) 조치와 경고도 포함될 수 있습니다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Field Safety Notice
  • 사례 연번
    SN2012(07)
  • 날짜
    2012-06-21
  • 사례 국가
  • 사례 출처
    HPRA
  • 사례 출처 URL
  • 비고 / 경고
    "Irish data is current through April 2019. All of the data comes from the Health Products Regulatory Authority (Ireland), except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of data from the U.S. and Ireland."
  • 데이터 추가 비고
  • 원인
    Abbott has determined that architect ca 19-9xr reagent lots 08849m500, 08851m500, 08852m500, 08853m500, 10122m500 and 10040m500 are demonstrating an upward shift in patient results. abbott has advised that through their investigations, a positive median bias of 83% has been observed when compared to other architect ca 19-9xr lots. the manufacturer advised that this issue may impact patient management when affected lots are used for monitoring patients.
  • 조치
    The IMB recommends that:  1. Customers consider the information in the FSN regarding the magnitude of the bias observed over the range of the assay (0-365%) and the need to review patient results.  2. Customers who determine the need to conduct a look back study and require further information in relation to this issue should contact Abbott for further advice. A copy of the FSN is also available on the HPRA website at the following link:  Urgent Field Safety Notice - V14968

Device

  • 모델명 / 제조번호(시리얼번호)
  • 제품 설명
    ARCHITECT CA 19-9XR
  • Manufacturer

Manufacturer

  • 제조사 모회사 (2017)
  • 제조사 의견
    “We are in constant communication with regulatory agencies and competent authorities worldwide which allows us to implement global recalls or in-country communication quickly and effectively,” Abbott, which now owns St. Jude Medical told ICIJ in a statement. In addition to sending global notices to physicians worldwide, we also make sure that product advisories are available online and classification of product recalls and product advisories are determined by global regulatory bodies which can impact the timing in any given country. MD companies follow varying regulations in different countries. In come countries software is not regulated so a recall in one country related to software would not be classified as a recall or field action in another. In addition, review cycles within the regulatory process can be different in each country which can impact communication and recall timing.
  • Source
    HPRA