Beckton Dickinson Vacutainer Multiple Sample Luer Adaptor 의 리콜

New Zealand Medicines and Medical Devices Safety Authority에 따르면, 해당 리콜 는 New Zealand 에서 BD Medical (USA) 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    14561
  • 사례 시작날짜
    2013-04-26
  • 사례 국가
  • 사례 출처
    NZMMDSA
  • 사례 출처 URL
  • 비고 / 경고
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • 데이터 추가 비고
    Recalling Organisation: Becton Dickinson Ltd, 14B George Bourke Drive, Mt Wellington, Auckland 1060
  • 원인
    Manufacturer has received a number of complaints relating to blood leakage when the bd vacutainer multile sample luer adapter is used in the blood collection process. these complaints relate to:, 1. the sleeve covering the np cannula, 2.Luer tip damage., due to current stocks holdings both locally and globally and existing demands bd is unable to substitute product for the bd vacutainer multiple sample luer adapter. should a defective device be used bd is providing information (product correction ) regarding the use of the device ., bd recommends that staff closley observe the luer tip of the device for potential damage. in the event that damage is observed staff are instructed to discard the individual device. should blood leak from the luer connection during blood collection ases the need to conduct a re-draw with a new luer adaptor, based upn the volume of blood in the tube.
  • 조치
    Manufacturer to issue advice regarding use

Device

  • 모델명 / 제조번호(시리얼번호)
    Model: 367300, Affected: all lot numbers manufactured starting with lot number 1354918 and after