Biomet 3i Dental Implants 의 리콜

New Zealand Medicines and Medical Devices Safety Authority에 따르면, 해당 리콜 는 New Zealand 에서 Implant Innovations Inc 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    17693
  • 사례 시작날짜
    2014-11-21
  • 사례 국가
  • 사례 출처
    NZMMDSA
  • 사례 출처 URL
  • 비고 / 경고
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • 데이터 추가 비고
    Recalling Organisation: Biomet 3i Australia Pty Ltd (NZ), 16 Huapai Street, Onehunga, Auckland 1061
  • 원인
    Biomet 3i has identified a visually observable condition which may be present on a limited number of the above identified implants. the condition is caused by the implant coming into contact with residual machining fluid which was not adequately degreased from a portion of 1 lot of packaging cylinders. the lot of packaging cylinders was used in the packaging of the 4 implant lots., the condition does not meet the cleanliness specification of biomet 3i. the residual machining fluid causes the implants to be discolored. although the residual machining fluid is not a health hazard, the discolored implant is visually unacceptable to customers and, thus, biomet 3i is removing these suspect implants from the market.
  • 조치
    Product to be returned to supplier

Device

  • 모델명 / 제조번호(시리얼번호)
    Model: Catalogue Numbers: BOST3211, Affected: Lots: 2014051395, 2014051817
  • Manufacturer

Manufacturer

  • 제조사 모회사 (2017)
  • 제조사 의견
    “The safety of patients has always been, and continues to be, Zimmer Biomet’s top priority and it is our honor to be a leader in this industry for the past 90 years,” Zimmer Biomet told ICIJ in a statement. “We adhere to strict regulatory standards, and work closely with the FDA and all applicable regulatory agencies in each of our regions as part of our commitment to operating a first-rate quality management system across our global manufacturing network. The company added that it is focused on staying at the forefront of innovation and doing right by the millions of patients who rely on the company’s products.
  • Source
    NZMMDSA