Brainlab Disposable Reflective Marker Spheres (DRMS) 의 리콜

New Zealand Medicines and Medical Devices Safety Authority에 따르면, 해당 리콜 는 New Zealand 에서 Northern Digital Inc 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    19094
  • 사례 시작날짜
    2015-11-09
  • 사례 국가
  • 사례 출처
    NZMMDSA
  • 사례 출처 URL
  • 비고 / 경고
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • 데이터 추가 비고
    Recalling Organisation: Johnson & Johnson (NZ) Limited, 13a Gabador Place, Mt Wellington, AUCKLAND
  • 원인
    Some drms may separate at the mid-point where the two halves of the sphere are sealed together. separation may occur during installation of the spheres onto reference arrays or surgical tools (i.E., when threading them onto posts), or potentially after installation on tools that are impacted with high forces. should such a drms separate during a surgical procedure, there is a possibility for:, - a portion of the sphere, its reflective coating and cured or uncured adhesive contacting patient tissues., - contamination of surgical trays, surgical drapes, surgical tools or physicians' gloves due to contact with the separated component., - prolongation of surgery due to the need to replace the broken sphere., the resulting separated sphere halves will not continue to be tracked by the brainlab image guided surgery system.
  • 조치
    Product to be returned to supplier

Device

  • 모델명 / 제조번호(시리얼번호)
    Model: 41773G, 41774G, 41772G, Affected: 1014213001, C103461502, C107761402, C100111501, C1034813001, C108341301, C101021502, C103881402, C108721402, C101161401, C104761402, C109051301, C101611502, C105541402, C109251402, C102311401, C106521402, C1098312001, C102481502, C106721301, C110411401