  • 사례 유형
    Recall
  • 사례 ID
    19733
  • 사례 시작날짜
    2016-02-05
  • 사례 국가
  • 사례 출처
    NZMMDSA
  • 사례 출처 URL
  • 비고 / 경고
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • 데이터 추가 비고
    Recalling Organisation: Abbott Vascular Division of Abbott Laboratories (NZ) Ltd, Ground Floor, Building D, 4 Pacific Rise, Mount Wellington, Auckland 1060
  • 원인
    Manufacturer has recently received reports of cases in clip delivery system devices that contain the one-way actuator knob where a user attempted implanting a mitralclip but the clip could not be detached from the delivery system due to a mandrel fracture. these cases resulted in severe surgical interventions and in one case the patient died post-operatively due to severe comorbidities., current instructions for use (ifu) require the arm positioner to be in a neutral position prior to turning the actuator knob to deploy the clip; manufacturerr is revising the mitraclip ifu clip deployment sequence to provide additional assurance that tension is completely eliminated prior to deploying the clip. abbott vascular will train all mitraclip implanters on the revised instructions., note : currently, there are no mitraclip systems in new zealand.
  • 조치
    Manufacturer to issue advice regarding use

Device

Manufacturer

  • 제조사 모회사 (2017)
  • 제조사 의견
    “We are in constant communication with regulatory agencies and competent authorities worldwide which allows us to implement global recalls or in-country communication quickly and effectively,” Abbott, which now owns St. Jude Medical told ICIJ in a statement. In addition to sending global notices to physicians worldwide, we also make sure that product advisories are available online and classification of product recalls and product advisories are determined by global regulatory bodies which can impact the timing in any given country. MD companies follow varying regulations in different countries. In come countries software is not regulated so a recall in one country related to software would not be classified as a recall or field action in another. In addition, review cycles within the regulatory process can be different in each country which can impact communication and recall timing.