GE X-Ray Systems with CRT Video Monitor Suspensions 의 리콜

New Zealand Medicines and Medical Devices Safety Authority에 따르면, 해당 리콜 는 New Zealand 에서 GE Medical Systems LLC 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    15819
  • 사례 시작날짜
    2013-12-09
  • 사례 국가
  • 사례 출처
    NZMMDSA
  • 사례 출처 URL
  • 비고 / 경고
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • 데이터 추가 비고
    Recalling Organisation: GE Healthcare Limited, 8 Tangihua Street, Auckland 1010
  • 원인
    It has been identified during several preventive maintenance activities that the 4 outer screws that hold the overhead video monitor suspension to the carriage were found missing since installation of the system. there is a potential that your system has not been installed per specified installation requirements. ge healthcare is not aware of any fall of the overhead monitor suspension; however, missing screws may increase the risk for fall of the monitor suspension. no injury has been reported due to this issue.
  • 조치
    Product to be modified

Device

  • 모델명 / 제조번호(시리얼번호)
    Model: Precision 500D, Legacy, Prestilix, Monitrol ,RFX, SFX, Tilt-C, Prestige VH, Prestige SI, and Prestige SII, Affected: All CRT Monitor suspension installed since Jan 1993 with products Precision 500D, Legacy, Prestilix, Monitrol ,RFX, SFX, Tilt-C, Prestige VH, Prestige SI, and Prestige SII
  • Manufacturer

Manufacturer