ID-DiaCell l-ll-lll ID-DiaPanel 의 리콜

New Zealand Medicines and Medical Devices Safety Authority에 따르면, 해당 리콜 는 New Zealand 에서 Bio-Rad Laboratories Inc 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    21117
  • 사례 시작날짜
    2017-02-28
  • 사례 국가
  • 사례 출처
    NZMMDSA
  • 사례 출처 URL
  • 비고 / 경고
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • 데이터 추가 비고
    Recalling Organisation: Bio-Rad Laboratories Ltd, 189 Bush Road, Rosedale, AUCKLAND 0632
  • 원인
    The manufacturer has become aware of unexpected reactions primarily on eluates and qc samples but also with some patient samples when using the id-diacell l-ll-lll and id-diapanel products. in case of doubtful reactions, please re-test with a new kit. if the results remain in doubt, the manufacturer advises to send the sample to a reference laboratory and if an urgent transfusion is required, perform a crossmatch., follow up 24 feb 2017: there has been further development (follow up) from our manufacturer additional 3 other devices that are also affected., the manufacturer received complaints regarding indirect antiglobulin test (iat), especially when eluates and qc samples were tested, less frequently when patient samples were tested.
  • 조치
    Manufacturer to issue advice regarding use

Device

  • 모델명 / 제조번호(시리얼번호)
    Model: , Affected: All lots currently in date
  • Manufacturer