Philips DigialDiagnost X-Ray System 의 리콜

New Zealand Medicines and Medical Devices Safety Authority에 따르면, 해당 리콜 는 New Zealand 에서 Philips Medical Systems DMC GmbH 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    13858
  • 사례 시작날짜
    2012-11-29
  • 사례 국가
  • 사례 출처
    NZMMDSA
  • 사례 출처 URL
  • 비고 / 경고
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • 데이터 추가 비고
    Recalling Organisation: Philips New Zealand Commercial Limited, Level 3, 123 Carlton Gore Road, Newmarket, AUCKLAND 1023
  • 원인
    With digitaldiagnost (didi) r 2.0.2 a mirrored "r in a circle" is burnt into the image (if mirrored) in the lower right corner of the image. when such an image is mirrored back inside the pacs system, an unmirrored "r in a circle" appears in the lower left corner. this can be mistaken for a "right patient side" marker, although this can appear on the left patient side., philips previously notified owners of affected digitaldiagnost systems that software release 2.1.3 would be installed on their systems to modify the appearance of the mirror icon, as described below. however, software release 2.1.3 contained a defect that may result in the incorrect placement of the electronic marker (left/right) indicating the patient side if an image is rotated.
  • 조치
    Software to be upgraded