Philips Heartstart MRx Monitor/Defibrillator 의 리콜

New Zealand Medicines and Medical Devices Safety Authority에 따르면, 해당 리콜 는 New Zealand 에서 Philips Medical Systems (USA) 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    14847
  • 사례 시작날짜
    2013-06-18
  • 사례 국가
  • 사례 출처
    NZMMDSA
  • 사례 출처 URL
  • 비고 / 경고
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • 데이터 추가 비고
    Recalling Organisation: Philips New Zealand Commercial Limited, Level 3, 123 Carlton Gore Road, Newmarket, AUCKLAND 1023
  • 원인
    The philips heartstart mrx monitor/defibrillator can deliver a non- synchronised shock in the synchronised cardioversion mode. this can occur if the user rotates the therapy knob while simultaneously pressing the sync button, then charges the mrx and presses the shock button., the mrx will still display r wave markers as if normally operating in synchronised cardioversion mode. the "sync" label will still be seen on the display. when the mrx is charged and the shock button pressed, the shock will be immediately delivered and may not be synchronised to the r wave., in addition, the label affixed to the top of the mrx device is not consistent with the mrx instructions for use on synchronised cardioversion.
  • 조치
    Software to be upgraded

Device

  • 모델명 / 제조번호(시리얼번호)
    Model: Models: M3535A, M3536A, M3536J, M3536M, M3536MC, M3536M2, M3536M4, M3536M5, M3536M6, Affected: Serial numbers within the range US00100100 to US00567299