Philips IntelliVue Information Centre (PIIC) iX 의 리콜

New Zealand Medicines and Medical Devices Safety Authority에 따르면, 해당 리콜 는 New Zealand 에서 Philips Medical Systems (USA) 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    18025
  • 사례 시작날짜
    2015-01-26
  • 사례 국가
  • 사례 출처
    NZMMDSA
  • 사례 출처 URL
  • 비고 / 경고
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • 데이터 추가 비고
    Recalling Organisation: Philips New Zealand Commercial Limited, Level 3, 123 Carlton Gore Road, Newmarket, AUCKLAND 1023
  • 원인
    Sp02 and/or non invasive blood pressure (nbp) alarms may become disabled without visual notification to the user., a software defect has been identified that involves the use of the "change bed label" feature at the information center ix. if the configuration for this item is set to "optional" and a user selects the "location" button within the "manage patient" menu and makes a change to the patient location, alarms for sp02 and/or non invasive blood pressure (nbp) will become disabled without visual notification (bell with an "x") to the user. this problem only happens when the patient is monitored using a networked intellivue trx telemetry m4841/trx4851a patient worn device (pwd).
  • 조치
    Software to be upgraded

Device

  • 모델명 / 제조번호(시리얼번호)
    Model: Part Numbers: 866023, 866024, 866117, Affected: , Software version: Revisions A.00, A.01, A.02
  • Manufacturer

Manufacturer

  • 제조사 모회사 (2017)
  • Source
    NZMMDSA