Physio-Control Lifepak 1000 Defibrillators 의 리콜

New Zealand Medicines and Medical Devices Safety Authority에 따르면, 해당 리콜 는 New Zealand 에서 Physio-Control Corporation 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    21045
  • 사례 시작날짜
    2017-02-03
  • 사례 국가
  • 사례 출처
    NZMMDSA
  • 사례 출처 URL
  • 비고 / 경고
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • 데이터 추가 비고
    Recalling Organisation: Medxus Ltd, Unit 4, 100 Fitgerald Avenue, Christchurch 8011
  • 원인
    Reports of incidents where customers have attempted to use their lifepak 1000 defibrillator and the device has shut down unexpectedly during patient treatment. this unexpected shut down is due to an intermittent connection between the battery and device contacts, the company has determined that this intermittent connection is a result of wear and subsequent oxidation formation between the battery and device electrical contacts. this condition can occur over time in lifepak 1000 devices that are exposed to vibration and have a battery installed for long periods of time. this issue can potentially affect any lifepak 1000 device, however customers with non-rechargeable batteries who do not routinely remove the battery for inspection, as indicated in the lifepak 1000 defibrillator operating instructions, are more susceptible to this issue.
  • 조치
    Manufacturer to issue advice regarding use