RaySearch RayStation Treatment Planning System 의 리콜

New Zealand Medicines and Medical Devices Safety Authority에 따르면, 해당 리콜 는 New Zealand 에서 RaySearch Laboratories AB 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    20245
  • 사례 시작날짜
    2016-06-07
  • 사례 국가
  • 사례 출처
    NZMMDSA
  • 사례 출처 URL
  • 비고 / 경고
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • 데이터 추가 비고
    Recalling Organisation: CARSL Consulting, 50B Puketapu Road, Taradale, Napier 4112
  • 원인
    An issue has been found with the treatment plan report in raystation 5. for a treatment plan with multiple beam sets, the table for roi plan dose statistics in the report may show the statistics for a beam set dose. the error can only be triggered when using a report template where statistics for a beam set dose are included as the last dose statistics prior to the plan dose statistics. for such a report template, all reports for multiple beam set plans would always be affected., the user must be aware of information to avoid incorrect assessment of treatment plan dose levels. i.E inspect all report templates that are intended to be used for plans with multiple beam sets. make sure that "poi dose statistics (plan)" is included immediately before "roi dose statistics [plan dose]" in the report content area., this issue will be resolved in the next version of raystation, scheduled for market release nov 2016.
  • 조치
    Manufacturer to issue advice regarding use

Device

  • 모델명 / 제조번호(시리얼번호)
    Model: , Affected: , Software version: 5.0.0.37, 5.0.1.11 and 5.0.2.35
  • Manufacturer

Manufacturer