S&N Birmingham Hip Resurfacing (BHR) Dysplasia Cup 의 리콜

New Zealand Medicines and Medical Devices Safety Authority에 따르면, 해당 리콜 는 New Zealand 에서 Smith & Nephew Orthopaedics Ltd 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    18028
  • 사례 시작날짜
    2015-02-05
  • 사례 국가
  • 사례 출처
    NZMMDSA
  • 사례 출처 URL
  • 비고 / 경고
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • 데이터 추가 비고
    Recalling Organisation: Smith & Nephew Surgical Ltd, Unit A 36 Hillside road, Wairau Valley, AUCKLAND 0760
  • 원인
    Smith & nephew has updated the instructions for use of the bhr system to include an expansion of the warnings for certain population sub-groups, in line with available data sources including registry data., if a patient from the following population sub-groups is implanted with a bhr system, the patient is at greater risk of requiring a revision surgery earlier than expected:, * females, * males aged 65 or greater, * patients requiring an implant head size < 48mm, patients who experience symptoms including pseudotumors, tissue masses, fluid collections, enlarges bursae, pain and swelling local buildup of excessive metal particles or metal hypersensitivity, may require revision surgery, with attendant risks and the potential for impaired function.
  • 조치
    Instructions for use to be updated

Device

  • 모델명 / 제조번호(시리얼번호)
    Model: 74120246, 74122248, 74120250, 74122252, 74120254, 74122256, 74120258, 74122260, 74120262, 74122264, 74120266, Affected: All lots
  • Manufacturer

Manufacturer