Siemens AXIOM Artis, Artis zee and Artis Q/Q 의 리콜

New Zealand Medicines and Medical Devices Safety Authority에 따르면, 해당 리콜 는 New Zealand 에서 Siemens AG 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    19817
  • 사례 시작날짜
    2016-03-01
  • 사례 국가
  • 사례 출처
    NZMMDSA
  • 사례 출처 URL
  • 비고 / 경고
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • 데이터 추가 비고
    Recalling Organisation: Siemens Healthcare (NZ) Ltd, Millennium Centre, Part Level 2, Building A, 600 Great South Road, Ellerslie, AUCKLAND 1051
  • 원인
    Due to a leak in the cooling system, fluids could seep into the equipment cabinet. this leak can occur sporadically in the affected systems., manufacturer initiated two field safety corrective actions (ax004/15/s und ax060/15/s) to address the issues with two different vendors for detector cooling devices. as the issues are identical, locally manufacturer sending both letters with one cover letter to all affected customers (this will prevent confusion among customers that have multiple systems and were required to receive both letters). each affected system will receive only one correction depending on the system type.
  • 조치
    Product to be modified

Device

  • 모델명 / 제조번호(시리얼번호)
    Model: 7555365,10094141,10094139,10094137,10094135,10280959,10848281,10848280,10848282,7413078,7412807,7008605,10848355,7728392,7555357,7727717, Affected:
  • Manufacturer

Manufacturer