St Jude Medical Ellipse DR CD2377-36 Implantable Defibrillator 의 리콜

New Zealand Medicines and Medical Devices Safety Authority에 따르면, 해당 리콜 는 New Zealand 에서 St Jude Medical (USA) 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    17202
  • 사례 시작날짜
    2014-08-19
  • 사례 국가
  • 사례 출처
    NZMMDSA
  • 사례 출처 URL
  • 비고 / 경고
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • 데이터 추가 비고
    Recalling Organisation: Abbott Medical New Zealand Limited, C/- Healthcare Logistics, 54 Carbine Road, Mt Wellington, Auckland 1641
  • 원인
    Extended charge time may present as a result of internal damage to the capacitors used in the high voltage charging circuitry of the subject devices., as designed, the device will deliver the available energy on the capacitors once the charge time limit of 32 seconds is reached, even if the energy is less than the programmed value., this anomaly may occur during capacitor maintenance or charging for high voltage therapy, and may result in delayed delivery of high voltage therapy and/or delivery of part of a programmed high voltage therapy shock., there have been no serious injuries or patient deaths reported to st. jude medical as a result of this anomaly.
  • 조치
    Manufacturer to issue advice regarding use

Device

  • 모델명 / 제조번호(시리얼번호)
    Model: Model: DR CD2377-36, Affected: Ellipse ICDs with serial numbers beginning with “1” and that are below 1131098, as well as Ellipse ICDs with serial numbers beginning with “7” that are below 7127327
  • Manufacturer

Manufacturer

  • 제조사 모회사 (2017)
  • 제조사 의견
    “We are in constant communication with regulatory agencies and competent authorities worldwide which allows us to implement global recalls or in-country communication quickly and effectively,” Abbott, which now owns St. Jude Medical told ICIJ in a statement. In addition to sending global notices to physicians worldwide, we also make sure that product advisories are available online and classification of product recalls and product advisories are determined by global regulatory bodies which can impact the timing in any given country. MD companies follow varying regulations in different countries. In come countries software is not regulated so a recall in one country related to software would not be classified as a recall or field action in another. In addition, review cycles within the regulatory process can be different in each country which can impact communication and recall timing.
  • Source
    NZMMDSA