  • 사례 유형
    Recall
  • 사례 ID
    17899
  • 사례 시작날짜
    2014-12-24
  • 사례 국가
  • 사례 출처
    NZMMDSA
  • 사례 출처 URL
  • 비고 / 경고
    Data from New Zealand is current through July 2018. All of the data comes from the New Zealand Medicines and Medical Devices Safety Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and New Zealand.
  • 데이터 추가 비고
    Recalling Organisation: Abbott Medical New Zealand Limited, C/- Healthcare Logistics, 54 Carbine Road, Mt Wellington, Auckland 1641
  • 원인
    Low incidence of back up operation in some implanted st. jude medical devices with radio frequency (rf) capability. this may occur as a result of a merlin@home transmitter initiating an implanted device software reset., this issue can only occur when the patient is being actively monitored by a merlin@home rf bedside transmitter.
  • 조치
    Software to be upgraded

Device

  • 모델명 / 제조번호(시리얼번호)
    Model: Model: EX1150, Affected: All units
  • Manufacturer

Manufacturer

  • 제조사 모회사 (2017)
  • 제조사 의견
    “We are in constant communication with regulatory agencies and competent authorities worldwide which allows us to implement global recalls or in-country communication quickly and effectively,” Abbott, which now owns St. Jude Medical told ICIJ in a statement. In addition to sending global notices to physicians worldwide, we also make sure that product advisories are available online and classification of product recalls and product advisories are determined by global regulatory bodies which can impact the timing in any given country. MD companies follow varying regulations in different countries. In come countries software is not regulated so a recall in one country related to software would not be classified as a recall or field action in another. In addition, review cycles within the regulatory process can be different in each country which can impact communication and recall timing.
  • Source
    NZMMDSA