translation missing: ko.activerecord.attributes.event.titles.recall_field_safety_notice

Saudi Food & Drug Authority에 따르면, 해당 translation missing: ko.activerecord.attributes.event.types.recall_field_safety_notice 는 Saudi Arabia 에서 Abbott 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

현장 안전성 서한은 의료기기 제조사 또는 제조사 대리인이 시중에 유통되고 있는 자사 의료기기 제품에 대하여 취할 수 있는 조치를 알리는 소통방식입니다. 주요 대상층은 보건의료업계 종사자와 의료기기 사용자들입니다. 여기에는 회수(recall) 조치와 경고도 포함될 수 있습니다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall / Field Safety Notice
  • 사례 ID
    mdprc 077 07 18 000
  • 사례 시작날짜
    2018-07-12
  • 사례 국가
  • 사례 출처
    SFDA
  • 사례 출처 URL
  • 비고 / 경고
    Saudi data is current through January 2019. All of the data comes from the Saudi Food & Drug Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Saudi Arabia.
  • 데이터 추가 비고
    Report Source: NCMDR
    SFDA urges all hospitals that have devices subjected to recall, to contact the company.
  • 원인
    Abbott has identified the following issues with the alinity ci‐series which may present a potential performance issue in alinity ci‐series software version 2.10. abbott is releasing alinity ci‐series software version 2.50 to correct these issues (see details in appendix a). 1. removing racks during the maintenance state is not detected by the system. issues with orders may occur when reloading a rack with samples and returning to running state. 2. orders created using more than one of the i‐series, c‐series, or calculated options to select assays will only include the assays selected in the last filtered view, if the order is saved prior to returning to the all option. 3. enabling an assay when the processing module is in the running or processing status causes the assay to be re‐installed. 4. a sample with 270 test orders remained in the scheduled status and would not process. 5. some maintenance and diagnostic procedures can appear to have completed successfully when there was insufficient inventory.

Device

  • 모델명 / 제조번호(시리얼번호)
    List Number (LN): 03R70‐01
  • 제품 설명
    Analyzers, Laboratory, Clinical Chemistry/Immunoassay
  • Manufacturer

Manufacturer

  • 제조사 모회사 (2017)
  • 제조사 의견
    “We are in constant communication with regulatory agencies and competent authorities worldwide which allows us to implement global recalls or in-country communication quickly and effectively,” Abbott, which now owns St. Jude Medical told ICIJ in a statement. In addition to sending global notices to physicians worldwide, we also make sure that product advisories are available online and classification of product recalls and product advisories are determined by global regulatory bodies which can impact the timing in any given country. MD companies follow varying regulations in different countries. In come countries software is not regulated so a recall in one country related to software would not be classified as a recall or field action in another. In addition, review cycles within the regulatory process can be different in each country which can impact communication and recall timing.
  • 제조사 대표
    Medical supplies & Services Co.Ltd Mediserv
  • Source
    SFDA