Saudi Food & Drug Authority에 따르면, 해당 translation missing: ko.activerecord.attributes.event.types.recall_field_safety_notice 는 Saudi Arabia 에서 Radiometer Medical 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

현장 안전성 서한은 의료기기 제조사 또는 제조사 대리인이 시중에 유통되고 있는 자사 의료기기 제품에 대하여 취할 수 있는 조치를 알리는 소통방식입니다. 주요 대상층은 보건의료업계 종사자와 의료기기 사용자들입니다. 여기에는 회수(recall) 조치와 경고도 포함될 수 있습니다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall / Field Safety Notice
  • 사례 ID
    mdprc 165 07 13 002
  • 사례 시작날짜
    2015-04-29
  • 사례 국가
  • 사례 출처
    SFDA
  • 사례 출처 URL
  • 비고 / 경고
    Saudi data is current through January 2019. All of the data comes from the Saudi Food & Drug Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Saudi Arabia.
  • 데이터 추가 비고
    Report Source: NCAR
    SFDA urges all hospitals that have devices subjected to recall, to contact the company
  • 원인
    Radiometer has become aware that some d788 pco2 membranes can cause biased measurement results on patient results as well as qc results. the bias can be seen after the replacement of the membrane, and will be positive in the low range and negative in the high range. there is a risk of wrong diagnosis and maltreatment of patients for biased measurement results in the high range , root cause investigation is still in progress, but the preliminary root cause is variations in the thickness of a silicone layer used in the membrane units. the error was discovered through internal testing on blood and qc material in connection with a capa raised to document the investigation of an issue regarding low qc comparison results. ....................... bfarm update action: your local radiometer representative has now installed the improved sealing ring for the ph electrode in your abl700 and abl800 series analyzers. the improved sealing ring may be identified by its blue color.

Device

  • 모델명 / 제조번호(시리얼번호)
    R338 to R493; part 942-063, pCO2, D788 Membrane ............................ Update code from MHRA: Select one of the Quality Control products listed below to be used for pCO2 verification: Qualicheck3+: S7330 (Level 1) Autocheck3+: S7335 (Level 1) Please see the attachment
  • 제품 설명
    Pco2 Membrane Box (D788), used on ABL700 and ABL800 series analysers
  • Manufacturer

Manufacturer

  • 제조사 모회사 (2017)
  • 제조사 대표
    Salehiya Trading Est., Riyadh, (01) 2628939
  • Source
    SFDA