  • 사례 유형
    Recall / Field Safety Notice
  • 사례 ID
    mdprc 092 12 18 000
  • 사례 시작날짜
    2018-12-20
  • 사례 국가
  • 사례 출처
    SFDA
  • 사례 출처 URL
  • 비고 / 경고
    Saudi data is current through January 2019. All of the data comes from the Saudi Food & Drug Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Saudi Arabia.
  • 데이터 추가 비고
    Report Source: NCMDR
    SFDA urges all hospitals that have devices subjected to recall, to contact the company.
  • 원인
    Abbott has identified that the incorrect components, 2 assay diluent bottles (7k61j) or 2 pretreatment 1 bottles (7k61u), may be present in a portion of batch 91345ui00. kits that do not contain all correct components cannot be scanned on the architect instrument, an error code will be generated upon loading on the reagent inventory screen (eg: error code 0900 or 0201).

Device

  • 모델명 / 제조번호(시리얼번호)
    List Number (LN): 7K61‐35 Lot Number: 91345UI00 UDI: N/A
  • 제품 설명
    IVD Test Reagent/Kits
  • Manufacturer

Manufacturer

  • 제조사 모회사 (2017)
  • 제조사 의견
    “We are in constant communication with regulatory agencies and competent authorities worldwide which allows us to implement global recalls or in-country communication quickly and effectively,” Abbott, which now owns St. Jude Medical told ICIJ in a statement. In addition to sending global notices to physicians worldwide, we also make sure that product advisories are available online and classification of product recalls and product advisories are determined by global regulatory bodies which can impact the timing in any given country. MD companies follow varying regulations in different countries. In come countries software is not regulated so a recall in one country related to software would not be classified as a recall or field action in another. In addition, review cycles within the regulatory process can be different in each country which can impact communication and recall timing.
  • 제조사 대표
    Medical supplies & Services Co.Ltd Mediserv
  • Source