Saudi Food & Drug Authority에 따르면, 해당 translation missing: ko.activerecord.attributes.event.types.recall_field_safety_notice 는 Saudi Arabia 에서 Baxter Healthcare 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

현장 안전성 서한은 의료기기 제조사 또는 제조사 대리인이 시중에 유통되고 있는 자사 의료기기 제품에 대하여 취할 수 있는 조치를 알리는 소통방식입니다. 주요 대상층은 보건의료업계 종사자와 의료기기 사용자들입니다. 여기에는 회수(recall) 조치와 경고도 포함될 수 있습니다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall / Field Safety Notice
  • 사례 ID
    mdprc 106 03 15 000
  • 사례 시작날짜
    2015-03-29
  • 사례 국가
  • 사례 출처
    SFDA
  • 사례 출처 URL
  • 비고 / 경고
    Saudi data is current through January 2019. All of the data comes from the Saudi Food & Drug Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Saudi Arabia.
  • 데이터 추가 비고
    Report Source: NCMDR
    SFDA urges all hospitals that have devices subjected to recall, to contact the company.
  • 원인
    Baxter healthcare is sending this communication to inform you of errors found in the arabic version of the instructions for use 'homechoice patient at home guide (pahg)' document number 071964016ara-sa, rev.C. the manual was intended to support saudi arabia's need to provide instructions for use on the following homechoice devices which have been distributed in saudi-arabia: - product codes r5c8320, 5c4474 - ce marked and for use with 230 v power supply (product code 5c4474 devices manufactured on or after 26 june 2013, serial numbers 650905 and onwards, were no longer ce marked). - product codes 5c4471, 5c8310 - non-ce marked and for use with 115 v power supply. the following errors were identified: 1) on the second page, it is not specified which product codes are ce marked and which codes are not. 2) on page 235, under electrical power requirements, the power requirements for the 115v configurations have been omitted. the correct supply voltage is mentioned on the labeling on the dataplate of the device, however, that label is in english and might not be understood by a patient using the device. 3) on the page 235 product codes r5c8320 and 5c8310 were incorrectly listed as "r5c8310" and "5c8320". a patient/user may erroneously plug a 115v device into a 230v outlet. this may cause the fuse within the homechoice device to blow rendering the device unusable. this may subsequently result in a slight delay in therapy as patients are able to continue therapy with manual supplies. most patients can also contact their pd centers or technical service for a swap within 24 hour.

Device

  • 모델명 / 제조번호(시리얼번호)
    Product Codes: 5C4471, 5C4474, 5C8310, R5C8320. (Only codes 5C831 0 and R5C8320 are being distributed is Saudi Arabia)
  • 제품 설명
    HomeChoice & HomeChoice Pro
  • Manufacturer

Manufacturer

  • 제조사 모회사 (2017)
  • 제조사 대표
    Arabian Health Care Supply Co. (AHCSC), Riyadh (01) 4747444
  • Source
    SFDA