translation missing: ko.activerecord.attributes.event.titles.recall_field_safety_notice

Saudi Food & Drug Authority에 따르면, 해당 translation missing: ko.activerecord.attributes.event.types.recall_field_safety_notice 는 Saudi Arabia 에서 Biomet 3i Inc 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

현장 안전성 서한은 의료기기 제조사 또는 제조사 대리인이 시중에 유통되고 있는 자사 의료기기 제품에 대하여 취할 수 있는 조치를 알리는 소통방식입니다. 주요 대상층은 보건의료업계 종사자와 의료기기 사용자들입니다. 여기에는 회수(recall) 조치와 경고도 포함될 수 있습니다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall / Field Safety Notice
  • 사례 ID
    mdprc 042 01 11 000
  • 사례 시작날짜
    2011-01-18
  • 사례 국가
  • 사례 출처
    SFDA
  • 사례 출처 URL
  • 비고 / 경고
    Saudi data is current through January 2019. All of the data comes from the Saudi Food & Drug Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Saudi Arabia.
  • 데이터 추가 비고
    Report Source: http://www.sfda.gov.sa/En/MedicalEquipments/recalls/NCAR+Brief.htm
    SFDA urges all hospitals that have devices subjected to recall, to contact the company.
  • 원인
    (january 17 , 2011 )the implant(s) may contain a manufacturing defect. this condition may be present in a small percentage of the lot 910695 and it relates to a shallow depth of the internal hex. the condition, if present, may prevent the driver; cover screw or abutment from fully engaging the implant. a complaint was received from a customer and the investigation showed that an incorrect machine tool was used during the manufacturing of the product and that related inspection protocols weren’t adequately followed.

Device

  • 모델명 / 제조번호(시리얼번호)
    OsseoTite Certain Implant 5X 11.5 mm Model IOSS511 , Lot/batch no. 910695
  • 제품 설명
    Dental implant endosseous
  • Manufacturer

Manufacturer

  • 제조사 모회사 (2017)
  • 제조사 의견
    “The safety of patients has always been, and continues to be, Zimmer Biomet’s top priority and it is our honor to be a leader in this industry for the past 90 years,” Zimmer Biomet told ICIJ in a statement. “We adhere to strict regulatory standards, and work closely with the FDA and all applicable regulatory agencies in each of our regions as part of our commitment to operating a first-rate quality management system across our global manufacturing network. The company added that it is focused on staying at the forefront of innovation and doing right by the millions of patients who rely on the company’s products.
  • 제조사 대표
    Asnan Medical Services , Riyadh , (01) 4621613
  • Source
    SFDA