  • 사례 유형
    Recall / Field Safety Notice
  • 사례 ID
    mdprc 023 01 15 000
  • 사례 시작날짜
    2015-01-05
  • 사례 국가
  • 사례 출처
    SFDA
  • 사례 출처 URL
  • 비고 / 경고
    Saudi data is current through January 2019. All of the data comes from the Saudi Food & Drug Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Saudi Arabia.
  • 데이터 추가 비고
    Report Source: NCMDR
    SFDA urges all hospitals that have devices subjected to recall, to contact the company.
  • 원인
    If the patient is connected to the anesthesia machine in a volume controlled ventilation mode, and attempts a large spontaneous breath, it is possible for the patient to create a high negative airway pressure around the same time as a volume controlled breath (mandatory breath).This pressure will be limited by the mechanical negative pressure limiter (npl) valve. the npl valve opens if the patient generates a negative pressure of -5 and -7.5 cmh20 (hpa). this opening pressure of the npl valve is defined at an inspiratory flow rate of 3.5-4.5 l/min. if the patient creates a higher flow rate than 4.5 l/min, the npl valve will continue to function but due to the resistance of the flow and time taken to reach the patient, the airway pressure may decrease further. the higher the flow generated by the patient, the greater the risk of increased negative airway pressure. if the patient is breathing spontaneously, support ventilation modes (vsv, psv) could be considered.

Device

  • 모델명 / 제조번호(시리얼번호)
    IntelliSave AX700 (P/N # 866205/10623-00) Dameca MRI 508 (P/N # 866203/10651MRI-00) Siesta i Whispa (P/N # 866202/10651-00) Siesta iTS (P/N # 866163/10653-00) Siesta Breasy (P/N # 866204/10652-00) All serial numbers are affected.
  • 제품 설명
    Anesthesia Machines
  • Manufacturer

Manufacturer

  • 제조사 모회사 (2017)
  • 제조사 대표
    Al-Faisaliah Medical System, Riyadh, (01) 2119948
  • Source
    SFDA