translation missing: ko.activerecord.attributes.event.titles.recall_field_safety_notice

Saudi Food & Drug Authority에 따르면, 해당 translation missing: ko.activerecord.attributes.event.types.recall_field_safety_notice 는 Saudi Arabia 에서 Advanced Surgical Design and Manufacture 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

현장 안전성 서한은 의료기기 제조사 또는 제조사 대리인이 시중에 유통되고 있는 자사 의료기기 제품에 대하여 취할 수 있는 조치를 알리는 소통방식입니다. 주요 대상층은 보건의료업계 종사자와 의료기기 사용자들입니다. 여기에는 회수(recall) 조치와 경고도 포함될 수 있습니다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall / Field Safety Notice
  • 사례 ID
    mdprc 050 12 12 000
  • 사례 시작날짜
    2012-12-09
  • 사례 국가
  • 사례 출처
    SFDA
  • 사례 출처 URL
  • 비고 / 경고
    Saudi data is current through January 2019. All of the data comes from the Saudi Food & Drug Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Saudi Arabia.
  • 데이터 추가 비고
    Report Source: NCAR
    SFDA urges all hospitals that have devices subjected to recall, to contact the company.
  • 원인
    Asdm is aware of elevated failure rates of the patella component of the active knee system, specifically related to the shearing of the pegs. electron microscopy of the failed patella has shown that the pegs are failing due to fatigue. the initiation site for the fatigue failure point appears to be a small overhang artefact, between 10 to 30 microns in depth, at the base of the patella peg. this small artefact may lead to elevated stresses at the base of the peg, and therefore increase the probability of premature product failure.

Device

  • 모델명 / 제조번호(시리얼번호)
    ARTG Number (ASDM Part Number, Lots Affected): 133900 (10-1514-075X, Lots manufactured prior to June 2012)
  • 제품 설명
    Prosthesis, internal, joint, knee, total - Patella
  • Manufacturer