Saudi Food & Drug Authority에 따르면, 해당 translation missing: ko.activerecord.attributes.event.types.recall_field_safety_notice 는 Saudi Arabia 에서 IMPAC Medical Systems Inc 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

현장 안전성 서한은 의료기기 제조사 또는 제조사 대리인이 시중에 유통되고 있는 자사 의료기기 제품에 대하여 취할 수 있는 조치를 알리는 소통방식입니다. 주요 대상층은 보건의료업계 종사자와 의료기기 사용자들입니다. 여기에는 회수(recall) 조치와 경고도 포함될 수 있습니다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall / Field Safety Notice
  • 사례 ID
    mdprc 115 07 13 000
  • 사례 시작날짜
    2013-07-17
  • 사례 국가
  • 사례 출처
    SFDA
  • 사례 출처 URL
  • 비고 / 경고
    Saudi data is current through January 2019. All of the data comes from the Saudi Food & Drug Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Saudi Arabia.
  • 데이터 추가 비고
    Report Source: NCMDR
    SFDA urges all hospitals that have devices subjected to recall, to contact the company.
  • 원인
    Sample couch parameter files are supplied with monaco releases 3.10.00 and later. the user can access these files from the treatment couch library. by clicking on a particular couch file, the individual components of the couch can be selected and their densities edited. the user can then use the “save as treatment couch” to enter a unique name for the modified couch parameter file. the sample files and their default values are not intended for clinical use, and monaco was designed with the assumption that the user would rename the file when it was populated with actual, clinical values. when sample couch parameter files are edited, the densities altered, and “save” is selected, the sample files will be updated with these new, user defined densities. subsequently, when a software upgrade is loaded, under the conditions described below, the modified couch densities in these sample files will be overwritten with the default value of 1.000. this problem will affect upgrades to monaco releases 3.20.00 and 3.30.00 from release 3.10.00 or 3.20.00. the problem will occur during an upgrade if and only if – prior to the upgrade – a sample couch parameter file was edited and saved under the original name. the problem occurs because the upgrade installation script is designed to overwrite the sample couch files in the treatment couch library with the original default values.

Device

  • 모델명 / 제조번호(시리얼번호)
    Versions 3.10.00 and later (2011-12-19 - Present)
  • 제품 설명
    Radiation therapy treatment planning system
  • Manufacturer

Manufacturer

  • 제조사 모회사 (2017)
  • 제조사 대표
    Analysis Device For Medical and Scientific Service ( ADMSS ), Dammam, (013) 8174422
  • Source