Saudi Food & Drug Authority에 따르면, 해당 translation missing: ko.activerecord.attributes.event.types.recall_field_safety_notice 는 Saudi Arabia 에서 ThermoFisher Scientific 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

현장 안전성 서한은 의료기기 제조사 또는 제조사 대리인이 시중에 유통되고 있는 자사 의료기기 제품에 대하여 취할 수 있는 조치를 알리는 소통방식입니다. 주요 대상층은 보건의료업계 종사자와 의료기기 사용자들입니다. 여기에는 회수(recall) 조치와 경고도 포함될 수 있습니다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall / Field Safety Notice
  • 사례 ID
    mdprc 001 05 12 000
  • 사례 시작날짜
    2012-05-01
  • 사례 국가
  • 사례 출처
    SFDA
  • 사례 출처 URL
  • 비고 / 경고
    Saudi data is current through January 2019. All of the data comes from the Saudi Food & Drug Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Saudi Arabia.
  • 데이터 추가 비고
    Report Source: NCMDR
    SFDA urges all hospitals that have devices subjected to recall, to contact the company.
  • 원인
    Whole blood samples (either fresh or frozen) for hba1c testing are pre-treated with multigent hemoglobin denaturant. the pre-treated samples are placed on the architect csystems and tested for glycated hemoglobin (hba1c) and total hemoglobin (thb) concentrations. according to the package insert, pretreated samples must be allowed to incubate for a minimum of 5 minutes at 15 to 30 °c prior to testing. pretreated samples are stated to be stable for “up to 8 hours at 15 to 30 °c and up to 48 hours if stored at 2 to 8 °c”. however, recent studies have demonstrated that hba1c measurements increase with time after the samples are pre-treated. therefore, it is necessary to change the handling instructions for pre-treating samples for the multigent hba1c assay. the investigation to determine root cause is ongoing.

Device

  • 모델명 / 제조번호(시리얼번호)
    CATALOG NO.: 02K96-20; LOT NO.: All lots within expiry date
  • 제품 설명
    Used in clinical laboratories for the quantitative in vitro measurement of percent HbA1c (HbA1c fraction) in human whole blood on the ARCHITECT cSystems and the AEROSETSystem. The HbA1c assay is intended to aid in the monitoring of long-term blood glucose control and compliance in individuals with diabetes mellitus.
  • Manufacturer

Manufacturer