translation missing: ko.activerecord.attributes.event.titles.recall_field_safety_notice

Saudi Food & Drug Authority에 따르면, 해당 translation missing: ko.activerecord.attributes.event.types.recall_field_safety_notice 는 Saudi Arabia 에서 Medtronic Inc 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

현장 안전성 서한은 의료기기 제조사 또는 제조사 대리인이 시중에 유통되고 있는 자사 의료기기 제품에 대하여 취할 수 있는 조치를 알리는 소통방식입니다. 주요 대상층은 보건의료업계 종사자와 의료기기 사용자들입니다. 여기에는 회수(recall) 조치와 경고도 포함될 수 있습니다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall / Field Safety Notice
  • 사례 ID
    mdprc 133 02 14 000
  • 사례 시작날짜
    2014-02-24
  • 사례 국가
  • 사례 출처
    SFDA
  • 사례 출처 URL
  • 비고 / 경고
    Saudi data is current through January 2019. All of the data comes from the Saudi Food & Drug Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Saudi Arabia.
  • 데이터 추가 비고
    Report Source: NCMDR
    SFDA urges all hospitals that have devices subjected to recall, to contact the company.
  • 원인
    The manufacturer has identified 14 reports over the last 30 years describing delayed spinal cord compression due to an epidural mass around a medtronic scs lead. all patients had neurological deficits and required surgical intervention to remove the mass. histologically, the epidural masses consisted of reactive tissue (granulomatous and/or fibrous tissue). the reported time to onset of neurological symptoms after lead implant ranged from weeks to 17 years. the severity ranged from muscle weakness to progressive quadriparesis. in several of the cases, only one side of the body was affected. the appearance of neurological symptoms was often preceded by a loss of therapy effect. the reported outcomes in these patients were full recovery (9 reports), permanent impairment (2 reports), and unknown (3 reports). there has not been a report of a patient death associated with this issue.

Device

  • 모델명 / 제조번호(시리얼번호)
    Details on the affected products is provided in the attachments
  • 제품 설명
    As an intervention for chronic back and/or leg pain, spinal cord stimulation can be an effective alternative or adjunct treatment to other therapies that have failed to manage pain on their own.
  • Manufacturer

Manufacturer

  • 제조사 모회사 (2017)
  • 제조사 의견
    “If our surveillance systems identify a potential performance issue, our personnel promptly evaluate the problem, including, when appropriate, conducting root cause investigations and internal testing to assess whether the product continues to meet specifications and defined performance criteria,” Medtronic told ICIJ in a statement. “In some cases, based on this evaluation, Medtronic may determine that a recall is necessary.” The company said that it communicates with healthcare providers and/or patients and provide recommendations to address such issues. Medtronic noted that these communications can include letters, emails, calls, press releases, physician notifications and social media postings, as well as informing the FDA and other regulators of the actions.
  • 제조사 대표
    Gulf Medical Co., (012) 6512828
  • Source
    SFDA