translation missing: ko.activerecord.attributes.event.titles.recall_field_safety_notice

Saudi Food & Drug Authority에 따르면, 해당 translation missing: ko.activerecord.attributes.event.types.recall_field_safety_notice 는 Saudi Arabia 에서 Stryker 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

현장 안전성 서한은 의료기기 제조사 또는 제조사 대리인이 시중에 유통되고 있는 자사 의료기기 제품에 대하여 취할 수 있는 조치를 알리는 소통방식입니다. 주요 대상층은 보건의료업계 종사자와 의료기기 사용자들입니다. 여기에는 회수(recall) 조치와 경고도 포함될 수 있습니다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall / Field Safety Notice
  • 사례 ID
    mdprc 028 11 15 000
  • 사례 시작날짜
    2015-11-08
  • 사례 국가
  • 사례 출처
    SFDA
  • 사례 출처 URL
  • 비고 / 경고
    Saudi data is current through January 2019. All of the data comes from the Saudi Food & Drug Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Saudi Arabia.
  • 데이터 추가 비고
    Report Source: NCMDR
    SFDA urges all hospitals that have devices subjected to recall, to contact the company.
  • 원인
    It was identified by a customer complaint that some i.V. poles were retracting/lowering from an extended position to the low height unexpectedly. an investigation was initiated which identified that the i.V. latch housing was nonconforming.

Device

  • 모델명 / 제조번호(시리얼번호)
    Manufacturer Model Number # 2035-112-010 Product Description : Head End Left, 2 Stage I.V. Pole Manufacturer Model Number # 2035-113-011 Product Description : Head End Right, 2 Stage I.V. Pole Manufacturer Model Number # 0785-035-101 Product Description : 26", 2 Stage IV Pole Assembly Manufacturer Model Number # 0785-035-200 Product Description : 26", 3 Stage IV Pole Assembly Manufacturer Model Number # 0785-035-300 Product Description : 30", 3 Stage IV Pole Assembly Manufacturer Model Number # 0785-035-401 Product Description : 30", 2 Stage IV Pole Assembly For I.V. Pole option numbers ship as service kits. ( Please see attachment )
  • 제품 설명
    Pop-up Push Handles, Surgical Accessory Rail
  • Manufacturer

Manufacturer

  • 제조사 모회사 (2017)
  • 제조사 대표
    Al-Faisaliah Medical System
  • Source
    SFDA