translation missing: ko.activerecord.attributes.event.titles.recall_field_safety_notice

Saudi Food & Drug Authority에 따르면, 해당 translation missing: ko.activerecord.attributes.event.types.recall_field_safety_notice 는 Saudi Arabia 에서 Elekta Inc 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

현장 안전성 서한은 의료기기 제조사 또는 제조사 대리인이 시중에 유통되고 있는 자사 의료기기 제품에 대하여 취할 수 있는 조치를 알리는 소통방식입니다. 주요 대상층은 보건의료업계 종사자와 의료기기 사용자들입니다. 여기에는 회수(recall) 조치와 경고도 포함될 수 있습니다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall / Field Safety Notice
  • 사례 ID
    mdprc 210 08 15 000
  • 사례 시작날짜
    2015-08-30
  • 사례 국가
  • 사례 출처
    SFDA
  • 사례 출처 URL
  • 비고 / 경고
    Saudi data is current through January 2019. All of the data comes from the Saudi Food & Drug Authority, except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Saudi Arabia.
  • 데이터 추가 비고
    Report Source: NCMDR
    SFDA urges all hospitals that have devices subjected to recall, to contact the company.
  • 원인
    Incorrect treatment delivery using third party fixed wedges. when fixed wedges are defined in source file maintenance, the user defines a maximum field size allowed for each wedge. this maximum field size is being ignored by the software when both fixed wedges and mlcs are present in a beam. the result is incorrect calculation and delivery of dose outside the physical extent of the wedge. the dose distribution calculated by xio will not match the dose delivered to the patient in regions beyond the physical dimension of the wedge. varian and siemens linacs will prevent delivery. elekta linacs with motorized wedges will prevent delivery. elekta linacs with third party fixed wedges will not prevent delivery.

Device

  • 모델명 / 제조번호(시리얼번호)
    Sites affected will be those: 1. Running XiO Release 4.51 and higher, and 2. Using Elekta linacs, and 3. Using third party fixed wedges in combination with MLCs
  • 제품 설명
    Radiation Treatment Planning System
  • Manufacturer

Manufacturer

  • 제조사 모회사 (2017)
  • 제조사 대표
    Analysis Device For Medical and Scientific Service ( ADMSS )
  • Source
    SFDA