Abbott Clinical Chemistry Phenobarbital 에 대한 현장 안전성 서한

ALIMS (Agencija za lekove i medicinska sredstva)에 따르면, 해당 현장 안전성 서한 는 Serbia 에서 Abbott Laboratories 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

현장 안전성 서한은 의료기기 제조사 또는 제조사 대리인이 시중에 유통되고 있는 자사 의료기기 제품에 대하여 취할 수 있는 조치를 알리는 소통방식입니다. 주요 대상층은 보건의료업계 종사자와 의료기기 사용자들입니다. 여기에는 회수(recall) 조치와 경고도 포함될 수 있습니다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Field Safety Notice
  • 날짜
    2012-11-16
  • 사례 출판 날짜
    2012-11-29
  • 사례 국가
  • 사례 출처
    ALIMSA
  • 사례 출처 URL
  • 비고 / 경고
    Serbian data is current through February 2018. All of the data comes from ALIMS (Agencija za lekove i medicinska sredstva), except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Serbia.
  • 데이터 추가 비고
  • 원인
    Abbott diagnostics withdraws phenobarbital reagents with numbers from the clinical chemistry market series 62299un12 and 85773un12 due to inaccuracy that increases with age of the reagent. the imprecision of the phenobarbital test is ≤ 7% of the total cv%; however, cv confirmed by internal testing was approximately 12% for all current reagent batches older than 5 months. the reagent with series number 52803un12, which has already expired, has also shown imprecision.

Device

  • 모델명 / 제조번호(시리얼번호)
    52803UN12, 62299UN12, 85773UN12
  • Manufacturer

Manufacturer

  • 제조사 모회사 (2017)
  • 제조사 의견
    “We are in constant communication with regulatory agencies and competent authorities worldwide which allows us to implement global recalls or in-country communication quickly and effectively,” Abbott, which now owns St. Jude Medical told ICIJ in a statement. In addition to sending global notices to physicians worldwide, we also make sure that product advisories are available online and classification of product recalls and product advisories are determined by global regulatory bodies which can impact the timing in any given country. MD companies follow varying regulations in different countries. In come countries software is not regulated so a recall in one country related to software would not be classified as a recall or field action in another. In addition, review cycles within the regulatory process can be different in each country which can impact communication and recall timing.
  • Source
    ALIMSA