SynchroMed lmplantable Infusion Pump 에 대한 현장 안전성 서한

ALIMS (Agencija za lekove i medicinska sredstva)에 따르면, 해당 현장 안전성 서한 는 Serbia 에서 Medtronic 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

현장 안전성 서한은 의료기기 제조사 또는 제조사 대리인이 시중에 유통되고 있는 자사 의료기기 제품에 대하여 취할 수 있는 조치를 알리는 소통방식입니다. 주요 대상층은 보건의료업계 종사자와 의료기기 사용자들입니다. 여기에는 회수(recall) 조치와 경고도 포함될 수 있습니다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Field Safety Notice
  • 사례 국가
  • 사례 출처
    ALIMSA
  • 비고 / 경고
    Serbian data is current through February 2018. All of the data comes from ALIMS (Agencija za lekove i medicinska sredstva), except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Serbia.
  • 데이터 추가 비고
  • 원인
    The synchromed priming bolus function is intended to quickly advance drug from the pump reservoir to the catheter tip to allow for therapy initiation while the patient remains under medical supervision. although drug is not intended to be delivered to the cerebrospinal fluid (csf) during the priming bolus, mixing of the drug and non-drug (sterile water/csf) fluids occurs at the high infusion rates used during a priming bolus. mixing results in the unintended delivery of drug prior to the end of the programmed bolus, as well as dilution of some of the drug remaining in the catheter at the end of the bolus. patients will receive unintended drug at a high rate of infusion in the csf during the priming bolus, and a period of reduced concentration of drug will occur following the priming bolus. .

Device

Manufacturer

  • 제조사 모회사 (2017)
  • 제조사 의견
    “If our surveillance systems identify a potential performance issue, our personnel promptly evaluate the problem, including, when appropriate, conducting root cause investigations and internal testing to assess whether the product continues to meet specifications and defined performance criteria,” Medtronic told ICIJ in a statement. “In some cases, based on this evaluation, Medtronic may determine that a recall is necessary.” The company said that it communicates with healthcare providers and/or patients and provide recommendations to address such issues. Medtronic noted that these communications can include letters, emails, calls, press releases, physician notifications and social media postings, as well as informing the FDA and other regulators of the actions.