200 mm sinus-SuperFlex-635 self-expanding nitinol stent system 에 대한 현장 안전성 서한

Agency for Medicinal Products and Medical Devices of the Republic of Slovenia (via FOI)에 따르면, 해당 현장 안전성 서한 는 Slovenia 에서 optimed Medizinische Instrumente GmbH 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

현장 안전성 서한은 의료기기 제조사 또는 제조사 대리인이 시중에 유통되고 있는 자사 의료기기 제품에 대하여 취할 수 있는 조치를 알리는 소통방식입니다. 주요 대상층은 보건의료업계 종사자와 의료기기 사용자들입니다. 여기에는 회수(recall) 조치와 경고도 포함될 수 있습니다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Field Safety Notice
  • 사례 국가
  • 사례 출처
    AMPMDRS
  • 비고 / 경고
    Data from Slovenia is current through February 2019. All of the data comes from the Agency for Medicinal Products and Medical Devices of the Republic of Slovenia (via FOI), except for the categories Manufacturer Parent Company and Product Classification.
    The Parent Company and the Product Classification were added by ICIJ.
    The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Slovenia.
  • 데이터 추가 비고
  • 원인
    This corrective action is based on the investigation of field complaints which indicates an increased number of deployment issues of sinus superflex-635 stents with 200 mm length. non-deployments were reported as well partial deployments of the stent. partially deployed stents may require an additional intervention by a doctor due to stent elongation or fracture. this recall does not affect any patient with an implanted stent of this length since this problem may only occur during the implantation of the device. .

Device

  • 모델명 / 제조번호(시리얼번호)
    8606-6200/ sinus-SuperFlex-635, Stent system, 6x200 mm, 6F/ 75 cm, .035i, application device 8606-7200/ sinus-SuperFlex-635, Stent system, 6x200 mm, 6F/ 120 cm, .035i, application device 8607-6200/ sinus-SuperFlex-635, Stent system, 7x200 mm, 6F/ 75 cm, .035i, application device 8607-7200/ sinus-SuperFlex-635, Stentsystem, 7x200 mm, 6F/ 120 cm, .035i, application device
  • Manufacturer

Manufacturer