Siemens healthcare diagnostics has confirmed that the advia centaur systems myoglobin
assay is not meeting the analytical sensitivity claim of < 3 ng/ml (ug/l) as specified in the
advia centaur myoglobin instructions for use (ifu), for all in-date reagent lots when evaluated
with calibrator u kit lots ending in 63 and 64. values up to 12 ng/ml (ug/l) were observed. .