Siemens healthcare diagnostics has confirmed that dimension® creatinine (cre2) assay
exhibits a negative bias at the low end of the urine analytical measurement range (amr) (see
table 2.). the limit of quantitation (loq) claim (5 mg/dl [442 µmol/l]) for urine samples is
not met. siemens is actively investigating the root cause of the issue and is working to
implement a solution. this issue also affects all future lots of cre2 until a solution is
implemented. the serum/plasma cre2 amr is not affected by this issue. .