Data from Slovenia is current through February 2019. All of the data comes from the Agency for Medicinal Products and Medical Devices of the Republic of Slovenia (via FOI), except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Slovenia.
데이터 추가 비고
원인
The reports are of inconsistent staining of cells and false positives reactivity with negative
control samples. the reports indicate the stain is not sufficiently removed from the adult cells
during the elution step; therefore it is difficult to distinguish fetal cells from the adult cells. a false
positive reaction with the negative control invalidates the test results and a repeat of the test is
needed. such repeat of testing may lead to delay in analysis of patient results and ultimately
delay in patient treatment. .