현장 안전성 서한은 의료기기 제조사 또는 제조사 대리인이 시중에 유통되고 있는 자사 의료기기 제품에 대하여 취할 수 있는 조치를 알리는 소통방식입니다. 주요 대상층은 보건의료업계 종사자와 의료기기 사용자들입니다. 여기에는 회수(recall) 조치와 경고도 포함될 수 있습니다.
Data from Slovenia is current through February 2019. All of the data comes from the Agency for Medicinal Products and Medical Devices of the Republic of Slovenia (via FOI), except for the categories Manufacturer Parent Company and Product Classification.
The Parent Company and the Product Classification were added by ICIJ.
The parent company information is based on 2017 public records. The device classification information comes from FDA’s Product Classification by Review Panel, based on matches of recall data from the U.S. and Slovenia.
데이터 추가 비고
원인
Trumpf medical has received notice from a supplier that it may eventually break due to a load in the spring arm welds when exposed to high forces. trumpf medical has not received any complaints or complaints about such failure. the supplier advises performing a visual inspection of the welds every two years as suggested in the july 2010 operating instructions. the inspection should therefore be performed for all operating lights, cameras and mounting systems (mounts) for trumpf medical monitors and for suspension arms.