Indicator, Biological Sterilization Process - Product Code FRC
원인
The 3m attest 1264/1264p biological indicators of eto sterilization contain a microbiological contaminant which can affect the performance of the positive control and the indicators to some limited extent. the color on a positive control may revert to negative after 24 hours of incubation.
조치
The firm conducted a field correction which began on 10/24/03. They provided revised labeling for the product. Letters, dated October 24, 2003, were sent to consignees. The letters had a description of the problem, instructions for doing the relabeling in the field, and recommendations for users of the product.