ARCHITECT Instrument 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Abbott Laboratories, Inc 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    78225
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-0165-2018
  • 사례 상황
    Open, Classified
  • 사례 국가
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    Analyzer, chemistry (photometric, discrete), for clinical use - Product Code JJE
  • 원인
    The peristaltic head tubing placed on the architect c4000, c8000, and c16000 has the possibility to leak due to a manufacturing issue associated with the tubing connector.
  • 조치
    All worldwide ARCHITECT Clinical Chemistry System customers with impacted instruments will receive the Product Correction Letter dated, 9/29/17. A customer reply is requested from all affected consignees.

Device

  • 모델명 / 제조번호(시리얼번호)
    c400073, c400103, c400498, c400600, c400890, c400934, c400935, c401018, c401055, c401056, c401182, c401505, c401520, c401686, c401887, c460196, c460200, c460943, c461550, c803669, c803753, c1600144, c1600249, c1600251, c1600641, c1600720, c1600819, c1600859, c1600976, c1601350, c1601351
  • 의료기기 분류등급
  • 의료기기 등급
    1
  • 이식된 장치?
    No
  • 유통
    Worldwide distribution to US, Canada, France, and Germany.
  • 제품 설명
    ARCHITECT c4000 part number 02P24; ARCHITECT c8000 part number 01G06; ARCHITECT c16000 part number 03L77 Clinical Chemistry systems.
  • Manufacturer

Manufacturer

  • 제조사 주소
    Abbott Laboratories, Inc, 1921 Hurd Dr, Irving TX 75038-4313
  • 제조사 모회사 (2017)
  • 제조사 의견
    “We are in constant communication with regulatory agencies and competent authorities worldwide which allows us to implement global recalls or in-country communication quickly and effectively,” Abbott, which now owns St. Jude Medical told ICIJ in a statement. In addition to sending global notices to physicians worldwide, we also make sure that product advisories are available online and classification of product recalls and product advisories are determined by global regulatory bodies which can impact the timing in any given country. MD companies follow varying regulations in different countries. In come countries software is not regulated so a recall in one country related to software would not be classified as a recall or field action in another. In addition, review cycles within the regulatory process can be different in each country which can impact communication and recall timing.
  • Source