U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
데이터 추가 비고
Non-Ac-Powered Patient Lift - Product Code FSA
원인
The pivot bolt that attaches the hanger bar to the scale of the patient lift can break, resulting in the hanger bar falling.
조치
Arjo sent an Urgent Device Recall Customer Notification letter dated 10/20/08 to all affected end users to the attention of the Administrator/Risk Manager via UPS 2nd day on the 10/20/08. The letters informed the accounts of the potential for the hanger bar to detach from the scale due to shearing of the pivot bolt. The accounts were advised to review the enclosed Safety Advisory Notice, note the affected model and serial numbers, and avoid using the lift under load on a high friction floor like carpet or soft tile while using the hanger bar to move the lift. The accounts were advised that an Arjo Field Service Technician would contact them within two weeks to schedule a visit to the account to perform repairs on their Maxi 500 lifts. Any questions were directed to Arjo Quality Department at 800-323-1245, ext. 6118. The accounts were also instructed to complete and fax back to Arjo the enclosed Customer Response Form, providing the facility name and address, contact name and phone number, the numbers of affected units, and acknowledgment of receipt and understanding of the Urgent Device Correction.
Arjo Maxi 500 Patient Lift with 2 Point Spreader Bar and Scale; a Non-AC-Powered Patient Lift; Model KM560101; BHM Medical Inc., Magog, QC, Canada || The Maxi 500 is a mobile passive lift intended to be for lifting and transferring of patients in hospitals, nursing homes, or other health care facilities.