AutoCheck5 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Radiometer America Inc 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    78097
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-3237-2017
  • 사례 시작날짜
    2017-09-07
  • 사례 상황
    Open, Classified
  • 사례 국가
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    Controls for blood-gases, (assayed and unassayed) - Product Code JJS
  • 원인
    Radiometer has recently become aware that a few of the autocheck 5+ level 2 ampoules, will give incorrect results for pco2 and po2. a defective qc ampoule will for: " pco2 give a result close to or just outside of the low end of the control range. (4.92  5.85 kpa or 36.9  46.9 mmhg) " po2 give a considerable higher result than the expected assigned value of 13.9 -14.3 kpa (104-107 mmhg).
  • 조치
    Radiometer sent a Safety Notice dated September 7, 2017, to all affected customers. The firm initiated their recall on 09/07/2017 by email. The letter requested the following: 1. Please check your inventory and analyzers for any boxes and ampoules of above mentioned QC. 2. Discard all Boxes and ampoules of this lot. 3. Complete the response form and return it to the number on the form. Please note: If you are not the end-user of the affected product, please ensure that this letter is distributed to the final end-user. Customers with questions were advised to contact their Radiometer representative.

Device

  • 모델명 / 제조번호(시리얼번호)
    S7745 Lot R0698
  • 의료기기 분류등급
  • 의료기기 등급
    1
  • 이식된 장치?
    No
  • 유통
    Worldwide Distribution - US (nationwide) Internationally to Canada, Netherlands, China, Czech Republic, Denmark, Germany, Spain France, Hungary, India, Denmark, Tokyo, South Korea, United Kingdom, Australia, Poland, South Africa, Singapore, Switzerland, Turkey, Norway, Austria, Italy, Sweden, and Finland.
  • 제품 설명
    AutoCheck5+, Level 2, specifically designed for Radiometer Americas blood gas analyzers || The QUALICHECK+ quality control (QC) ampoules are specifically designed for Radiometer America s blood gas analyzers
  • Manufacturer

Manufacturer

  • 제조사 주소
    Radiometer America Inc, 250 S Kraemer Blvd, Brea CA 92821-6232
  • 제조사 모회사 (2017)
  • Source
    USFDA