AutolubeIII 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 The Anspach Effort, Inc. 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    76059
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-1934-2017
  • 사례 시작날짜
    2016-10-06
  • 사례 상황
    Open, Classified
  • 사례 국가
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    Motor, surgical instrument, ac-powered - Product Code GEY
  • 원인
    Supplied directions for use (dfu) did not contain a recommended service interval.
  • 조치
    Consignees were sent a recall notice via mail, dated 10/10/2016. In the notice, the firm notifies consignees of the Directions For Use (DFU) correction that will include recommended servicing intervals. All DFUs distributed after 09/28/2016 will have the correct DFU. The recalling firm instructs consignees to review the service intervals provided for product received, forward the notification to necessary parties and/or facilities, and complete the attached Verification Form. Complete Verification Forms are to be sent by fax to 877-546-5069 or by email to Anspach4870@stericycle.com. Consignees are also asked to maintain a copy of the notice for their records. Any questions can be directed to Customer Service at 1-800-327-6887, option 1.

Device

  • 모델명 / 제조번호(시리얼번호)
    All product shipped prior to 09/27/2016;  DFU 18-0040; Part No. AUTOLUBE-III
  • 의료기기 분류등급
  • 의료기기 등급
    1
  • 이식된 장치?
    No
  • 유통
    Worldwide Distribution - US Nationwide in the states of HI, DC including PR and the countries of Guam and Japan
  • 제품 설명
    Autolube-III || Product Usage: || Pneumatic system
  • Manufacturer

Manufacturer

  • 제조사 주소
    The Anspach Effort, Inc., 4500 Riverside Dr, Palm Beach Gardens FL 33410-4235
  • 제조사 모회사 (2017)
  • Source
    USFDA