U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
The Parent Company was added by ICIJ.
The parent company information is based on 2017 public records.
데이터 추가 비고
Culture bottle (anaerobic) - Product Code MDB
원인
Recovery compromised-during manufacturing ambient air (including oxygen) was inadvertently introduced into the culture bottles, changing their reduction/oxidation potential thus leading to a compromised recovery of some obligate anaerobes and longer detection times.
조치
Affected Customers were notified of this recall on August 18, 2006 and August 21, 2006 via phone and instructed to destroy product remaining in stock.
Nationwide Distribution ---- including states of AL, CA, CT, FL, GA, IA, IL, IN, KY, LA, MA, MD, MI, MS, MT, NC, NE, NJ, NY, OH, PA, RI, TX, VA, WA, WI, and WV.
제품 설명
bioMerieux BacT/ALERT SN Anaerobic Culture Bottle-REF 259790