Baxter Interlink System MicroInfusion Manifold 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Baxter Healthcare Corp. 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    56892
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-0502-2011
  • 사례 시작날짜
    2010-10-21
  • 사례 출판 날짜
    2010-12-01
  • 사례 상황
    Terminated
  • 사례 국가
  • 사례 종료 날짜
    2011-10-16
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    Intravascular Administration Set - Product Code FPA
  • 원인
    If the manifolds are connected in series, there may be a risk of the manifolds loosening or disconnecting at the luer connection between them.
  • 조치
    Baxter sent an "Urgent Device Correction" letter dated October 20, 2010, via first class mail to the direct accounts, to the attention of the Director of Nursing and the Director of Materials Management, informing them that there may be a risk of the manifolds loosening or disconnecting at the luer connection between them. The loosening or disconnection can result in an interruption of therapy, potentially resulting in patient injury and death. The accounts were requested to place the notification where the inventory of affected product is dispensed, so that users are made aware of the procedures to follow: "If these devices are connected in series, and the luer connector is tightened, do not twist the manifolds. This may lead to loosening or disconnection, which may cause a leak. Additionally, Baxter recommends checking all connections before starting and during infusion therapy." The accounts were requested to complete the attached customer reply form confirming their receipt of the letter and fax it to Baxter at the number provided on the form. Dealers, wholesalers and distributors were requested to notify their customers of this action so they can comply with the procedures listed above. Any questions regarding the communication were directed to The Center for One Baxter at 1-800-422-9837 during the hours of 8:00 am to 5:00 pm CST.

Device

  • 모델명 / 제조번호(시리얼번호)
    product code 2N3410 and A2N3410 , all lots
  • 의료기기 분류등급
  • 의료기기 등급
    2
  • 이식된 장치?
    No
  • 유통
    Worldwide Distribution -- United States and Japan.
  • 제품 설명
    Baxter Interlink System, Micro-Infusion Manifold; and Interlink Manifold. An Rx sterile device with 3 injection sites, volume per injection site 0.1 mL, priming volume 0.8 mL, length 8.7 cm. Product code 2N3410 and A2N3410 - Japan || Baxter Healthcare corporation, Deerfield, IL 60015 USA; || Intended for use with a vascular access device for blood sampling and the administration of multiple drugs and solutions.