BD¿ PhoenixSpec¿ Calibrator Kit, consisting of 3 tubes identified as 0.25, 1.0 and 4.0 McFarland unit calibrators, and labeled in part ***Becton, Dickinson and Company, 7 Loveton Circle, Sparks, MD 21152 USA*** 의 리콜

U.S. Food and Drug Administration에 따르면, 해당 리콜 는 United States 에서 Becton Dickinson & Co. 에 의해 제조된 제품과 관련되어 있습니다.

이것은 무엇인가요?

의료기기에 문제가 생겼을 경우 제조사가 이를 바로잡거나 시장에서 회수하는 조치를 말한다. 회수(Recall)는 의료기기에 결함이 있거나, 건강에 위협이 되거나, 또는 결함도 있고 건강에도 위협이 될 경우에 발생한다.

데이터에 대해 더 자세히 알아보기 여기
  • 사례 유형
    Recall
  • 사례 ID
    34202
  • 사례 위험등급
    Class 2
  • 사례 연번
    Z-0430-06
  • 사례 시작날짜
    2005-12-01
  • 사례 출판 날짜
    2006-01-24
  • 사례 상황
    Terminated
  • 사례 국가
  • 사례 종료 날짜
    2006-04-06
  • 사례 출처
    USFDA
  • 사례 출처 URL
  • 비고 / 경고
    U.S. data is current through June 2018. All of the data comes from the U.S. Food and Drug Administration, except for the category Manufacturer Parent Company.
    The Parent Company was added by ICIJ.
    The parent company information is based on 2017 public records.
  • 데이터 추가 비고
    Nephelometer, For Clinical Use - Product Code JQX
  • 원인
    A calibrator for an in vitro diagnostic kit is not standarized correctly and may cause a calibration error when measuring microbiological cultures for turbidity.
  • 조치
    The recalling firm notified consignees by phone, fax and e-mail on 12/01/05. The notification advised that nephelometers displaying 'CAL' error messages on the LED should not be used until calibrated with a different kit lot. Consignees were requested to discard referenced lots for replacement. A response form was included in the notification.

Device

  • 모델명 / 제조번호(시리얼번호)
    Lot 5075281 Exp 2/14/06*** Lot 5115146 Exp 2/14/06
  • 의료기기 분류등급
  • 의료기기 등급
    1
  • 이식된 장치?
    No
  • 유통
    Product was sold to direct accounts nationwide and international affiliated distributors in Japan, Belize, Taiwan, Canada, Korea, Australia, and the Philippines.
  • 제품 설명
    BD¿ PhoenixSpec¿ Calibrator Kit, consisting of 3 tubes identified as 0.25, 1.0 and 4.0 McFarland unit calibrators, and labeled in part ***Becton, Dickinson and Company, 7 Loveton Circle, Sparks, MD 21152 USA***